1. |
Equipment Description |
1.1 |
Supplied by AWTA Ltd for sales within Australia & New Zealand: |
|
Sirolan Laserscan instrument |
|
|
IBM compatible PC |
- The operating system for the PC MUST be Windows XP.
|
|
|
- Includes new PCI analogue to digital conversion card interface between the PC and the Laserscan instrument.
|
|
Laserscan software |
- Windows software AWIN Ver 1.2.14 running under Windows XP.
|
|
|
- Includes new auto-dumping feature which empties the mixing basin automatically at the end of the measurement cycle. This can be turned on or off.
|
|
|
- Includes new solvent calculator to assist in adjusting the water content of the isopropanol solvent to within the specified limits
|
|
|
- Includes new CALHELPER utility to assist in calibrating the instrument.
|
|
Operator's manual |
|
|
Samples of 8 Interwoollabs calibration tops |
|
|
Accessories kit |
Includes: tweezers, hydrometer, clamps, thermometer, wash bottle, etc |
|
|
|
1.2 |
Other Equipment Items |
|
Essential items: |
- A supply of isopropanol solvent used as the carrier liquid (approx. 4 litres)
|
|
|
- A supply of distilled water (approx. 1 litre)
|
|
|
- Conditioned room with standard conditions of 20 degrees centigrade and relative humidity of 65% (for IWTO-12 test method)
|
|
Recommended items: |
- Minicore for preparing snippets from scoured or greasy wool samples.
|
|
|
- Guillotine for preparing snippets from top samples.
|
|
|
- Snippet Forced Air Drier for cleaning and drying greasy wool samples.
|
|
|
|
1.3 |
Services: |
|
|
Power supply |
- 10 Amp, 50 HZ 115 Volt or 240 Volt
|
|
|
|
1.4 |
Dimensions: |
|
|
Packages |
- 1 off Laserscan, 135kg, 100x100x100 cms
|
|
|
- 1 off PC, 15.5kg, 70x70x30 cms
|
|
|
- 1 off PC monitor, 18.5kg, 50x50x50 cms
|
|
Total Weight |
|
|
|
|
1.5 |
Recommended Operating Conditions and Training: |
|
Temperature: |
- 16 to 24 degrees centigrade.
|
|
Humidity: |
- Between 20% and 80% relative humidity (non-condensing).
|
|
Electrical noise: |
- Ensure there are no nearby sources of electrical noise (large electric motors, welding equipment, etc).
|
|
Mechanical shocks: |
- The Laserscan must be operated on a stable vibration free floor.
- The instrument is classified as "very delicate" and falls into the transport category of medical equipment. For example, it must be transported with an optical bench restraining strap deployed. Special care should be taken when transporting the instrument on unsealed roads, for example air cushions.
|
|
Dust: |
- A dust free environment is highly recommended.
|
|
Ventilation: |
- The rear of the instrument should not be adjacent to a wall.
|
|
Mains Power: |
- Supply voltage must be within the range of 220 and 240 VAC.
- A generator should be checked for compliance with this range by a licensed electrician prior to use.
|
|
Training: |
- All customers are required to have 1 day of basic operating training, and should be appropriately qualified or have adequate technical support facilities.
|
|
|
|