LASERSCAN

The Sirolan-LaserscanTM is used to determine: the Mean Fibre Diameter, distribution of diameter (SD), Coefficient of Variation (CV), % of fibres less than 30 micron (Comfort Factor), distribution histogram, and Fibre Curvature of clean wool conditioned in a standard atmosphere. The relevant test method is IWTO-12.

The measurements are carried out by passing prepared fibre snippets, which are automatically dispersed in a carrier liquid, through a measurement cell. In the cell, fibres intercept a laser beam and the resulting changes in beam intensity are converted to diameter readings by a computer. A parallel optoelectronic processing circuit and fibre optic discriminator is used to confirm valid measurements and reject non-fibre particles.

The instrument is calibrated using 8 standard wool tops.

General Features

Item

Inclusions

Additional Comments

Sirolan™ Laserscan

1x Sirolan™ Laserscan Instrument

Measures the fineness characteristics of wool fibres.

Click here to view larger image

View Larger Image

Click here to view larger image
View Larger Image

1x IBM Compatible PC

1x Laserscan Software

1x Operating Manual

English language only.

8x Interwoolabs calibration tops

Samples of the current series of Interwoollabs tops are supplied for calibrating the instrument.

1x Accessories kit

The accessories kit includes items for routine maintenance of the instrument.

 

Specification

1.

Equipment Description

1.1

Supplied by AWTA Ltd for sales within Australia & New Zealand:

Sirolan Laserscan instrument

  • CE Mark compatible

IBM compatible PC

  • The operating system for the PC MUST be Windows XP.
  • Includes new PCI analogue to digital conversion card interface between the PC and the Laserscan instrument.

Laserscan software

  • Windows software AWIN Ver 1.2.14 running under Windows XP.
  • Includes new auto-dumping feature which empties the mixing basin automatically at the end of the measurement cycle. This can be turned on or off.
  • Includes new solvent calculator to assist in adjusting the water content of the isopropanol solvent to within the specified limits
  • Includes new CALHELPER utility to assist in calibrating the instrument.

Operator's manual

Samples of 8 Interwoollabs calibration tops

Accessories kit 

Includes: tweezers, hydrometer, clamps, thermometer, wash bottle, etc

1.2

Other Equipment Items

Essential items:

  • A supply of isopropanol solvent used as the carrier liquid (approx. 4 litres)
  • A supply of distilled water (approx. 1 litre)
  • Conditioned room with standard conditions of 20 degrees centigrade and relative humidity of 65% (for IWTO-12 test method)

Recommended items:

  • Minicore for preparing snippets from scoured or greasy wool samples.
  • Guillotine for preparing snippets from top samples.
  • Snippet Forced Air Drier for cleaning and drying greasy wool samples.
  • A computer printer.

1.3

Services:

Power supply

  • 10 Amp, 50 HZ 115 Volt or 240 Volt

1.4

Dimensions:

Packages

  • 1 off Laserscan, 135kg, 100x100x100 cms
  • 1 off PC, 15.5kg, 70x70x30 cms
  • 1 off PC monitor, 18.5kg, 50x50x50 cms

Total Weight

  • 169 kgs

1.5

Recommended Operating Conditions and Training:

Temperature:

  • 16 to 24 degrees centigrade.

Humidity:

  • Between 20% and 80% relative humidity (non-condensing).

Electrical noise:

  • Ensure there are no nearby sources of electrical noise (large electric motors, welding equipment, etc).

Mechanical shocks:

  • The Laserscan must be operated on a stable vibration free floor.
  • The instrument is classified as "very delicate" and falls into the transport category of medical equipment. For example, it must be transported with an optical bench restraining strap deployed. Special care should be taken when transporting the instrument on unsealed roads, for example air cushions.

Dust:

  • A dust free environment is highly recommended.

Ventilation:

  • The rear of the instrument should not be adjacent to a wall.

Mains Power:

  • Supply voltage must be within the range of 220 and 240 VAC.
  • A generator should be checked for compliance with this range by a licensed electrician prior to use.

Training:

  • All customers are required to have 1 day of basic operating training, and should be appropriately qualified or have adequate technical support facilities.